Calibration is required in the process of controlling our equipment if we are to ensure the traceability of measurements.

With the help of etalon weights we verify the accuracy of our indication, eccentricity and repetability.
Calibration laboratory SA LK-008
![]() | Annex to the Accreditation Certificate | LK-008 |
| Issued on | 15. junij 2011 | |
| Replaces Annex dated | 1. april 2011 | |
| Information on current accreditation status is available at: | www.sa.gov.si |
| In laboratory |
| Measured quantity | Range | Calibration and measuring capability CMC (k=2)* | Notes |
| 3 |
| TIME AND FREQUENCY |
| 3.1.1 |
| Time interval |
| Timers |
| Time | 24h | 0,1s | By comparison In laboratory and on-site |
| 3.2.3 |
| Tachometers |
| Equipment for tachographs inspection |
| Speed Travelled distance Time interval | 20 km/h to 180 km/h 1000 m to 10000 m 24h | 0,2 % 0,2 % 2s / 24h | By comparison In laboratory and on-site |
| Speedometers |
| 10 to 70 km/h | 0,09 km/h | By comparison In laboratory and on-site |
| 4 |
| DIMENSIONAL QUANTITIES |
| 4.1 |
| Length |
| 4.1.2 |
| Length measuring devices |
| Guage block | 0,5 mm to 10 mm nad 10 mm to 125 mm nad 125 mm to 1000 mm | 70 nm 60 nm + 1· 10-6·L 100 nm + 1,6·10-6·L | By comparison | |
| Step gauges | 0 mm to 1000 mm | 1,25 μm + 5,5·10-5·L | Measuring area: < 100 mm x 100 mm | |
| Dial gauges | 0 mm to 100 mm | 0,4 μm + 2·10-5·L | In laboratory and on-site. On-site calibration only with load | |
| Tire depth gauges | 0 mm to 100 mm | 0,03 mm | In laboratory and on-site | |
| Electronic probes | 0 mm to 100 mm | 0,4 μm + 2·10-5·L | In laboratory and on-site Calibration on-site without evaluation of hysteresis | |
| Thickness gauges | 0 mm to 1000 mm | 0,2 μm + 6·10-6·L | Measuring area: < 100 mm x 100 mm | |
| Gap gauges | 0 mm to 1000 mm | 1,25 μm + 5,5·10-6·L | Measuring area: < 100 mm x 100 mm |
| 4.1.3 |
| Line scales, distances |
| Rulers/td> | 0 mm to 4000 mm | 50 μm + 3·10-6·L | ||
| Tape measures | 0 mm to 200 m | 50 μm + 3·10-6·L |
| 4.1.4 |
| Length measuring devices |
| Two point micrometers (external and internal) | 0 mm to 2000 mm | 1,5 μm + 5·10-6·L | ||
| Vernier calliper gauges | 0 mm to 2000 mm | 7,5 μm + 8·10-6·L | ||
| Laser distance meters | 0 mm to 4000 mm | 0,1 mm | ||
| Ultrasonic thickness meters | 0 mm to 50 mm | 0,15 mm + 6·10-3·L (L v mm) | ||
| Magnetic thickness meters | 0 mm to 1500 μm | 7 μm | ||
| Distance measuring wheels | 0 m to 1000 m | 10 mm + 0,15·L | ||
| Test sieves | to 400 mm | 3,8 μm + 5·10-6·L | ISO 3310 |
| 4.1.5 |
| Diameter |
| Internal diameter |
| Three point micrometers | 0 mm to 300 mm | 1,5 μm + 5·10-6·L | ||
| Ring gauges | 2 mm to 800 mm | 1,25 μm + 5,5·10-6·L |
| External diameter |
| Plug gauges | 0 mm to 300 mm | 0,2 μm + 6·10-6·L |
| 4.1.6 |
| Form error |
| Surface plates | to 500 mm x 700 mm | 3 μm | ||
| Rulers (flatness and parallelism) | to 1000 mm | 3 μm |
| 4.1.8 |
| Thread quantities |
| Screw ring gauges |
| Medium diameter | 3 mm to 300 mm | α=30°; 5 μm + 5·10-5·L α=55°; 3 μm + 7,5·10-6·L α=60°; 3,5 μm + 6·10-6·L | Rate in the range 0,35 mm to 6mm α= thread angle (according to EURAMET/cg.10/v.01, method 1a) |
| Screw plug gauges |
| Medium diameter | 3 mm to 300 mm | α=30°; 5 μm + 5·10-5·L α=55°; 3 μm + 7,5·10-6·L α=60°; 3,5 μm + 6·10-6·L | Rate in the range 0,35 mm to 6mm α= thread angle (according to EURAMET/cg.10/v.01, method 1a) |
| 4.1.9 |
| Co-ordinate measuring machines |
| 1D machines | 0 mm to 1000 mm | 0,2 μm + 4·10-6·L | In laboratory and on-site | |
| 2D machines | to 200 mm x 300 mm | 1,5 μm + 3·10-6·L | In laboratory and on-site | |
| 3D machines | to 1500 mm x 1500 mm | 1,0 μm + 2,5·10-6·L | In laboratory and on-site |
| 4.2 |
| Angle |
| 4.2.1 |
| Angle gauges |
| Squares | to 500 mm x 1000 mm | 6˝ |
| 4.2.3 |
| Clinometers |
| Protractor | 0° to 360° | 15˝ | ||
| level meters | -5 mm to 5 mm -0,8° to 0,8° | 25 µm/m 5˝ |
| 5 |
| MECHANICAL QUANTITIES |
| 5.1 |
| Force |
| Force transducers |
| > 1 N to 10 N 10 N to 100 N 100 N to 200 N 200 N to 400 N 400 N to 600 N 600 N to 1000 N 1000 N to 1500 N 1500 N to 2000 N 2000 N to 2500 N 2500 N to 5000 N | 0,011 N 0,11 N 0,11 N 0,11 N 0,12 N 0,13 N 1,1 N 1,1 N 1,2 N 3,5 N | By comparison In laboratory and on-site |
| Levers for calibration of roll brake testers |
| to 0,2 kN 0,2 kN to 0,5 kN 0,5 KN to 2,0 kN | 0,7 % 0,5 % 0,4 % | Levers for truck roll brake testers | ||
| 0,2 kN to 20 kN | 0,4 % | Levers for truck roll brake testers |
| 5.2 |
| Conventional mass |
| Etalon weights |
| 1 mg 2 mg 5 mg 10 mg 20 mg 50 mg 100 mg 200 mg 500 mg 1 g 2 g 5 g 10 g 20 g 50 g 100 g 200 g 500 g 1 kg 2 kg 5 kg 10 kg 20 kg 50 kg | 0,002 mg 0,002 mg 0,002 mg 0,002 mg 0,003 mg 0,004 mg 0,005 mg 0,006 mg 0,008 mg 0,010 mg 0,012 mg 0,015 mg 0,020 mg 0,025 mg 0,030 mg 0,05 mg 0,10 mg 0,25 mg 1,5 mg 3,0 mg 7,5 mg 15 mg 30 mg 250 mg |
| Objects of other optional masses |
| to 0,5 g 0,5 g to 5,1 g 5,1g to 50 g 50 g to 205 g 205 g to 2 kg 2 kg to 5 kg 5 kg to 10,1 kg 10,1 kg to 25 kg 25 kg to 40 kg 40 kg to 64,1 kg | 0,01 mg 0,03 mg 0,05 mg 0,1 mg 3,0 mg 5,0 mg 7,0 mg 25 mg 80 mg 120 mg |
| 5.3 |
| Scales |
| Non-automatic scales |
| to 5 mg 5 mg to 10 mg 10 mg to 20 mg 20 mg to 50 mg 50 mg to 100 mg 100 mg to 200 mg 200 mg to 500 mg 0,5 g to 1 g 1 g to 2 g 2 g to 5 g 5 g to 10 g 10 g to 20 g 20 g to 50 g 50 g to 600 g 600 g to 25 kg 25 kg to 100 kg 100 kg to 600 kg | 0,0013 mg 0,0015 mg 0,0016 mg 0,0020 mg 0,0024 mg 0,0030 mg 0,0039 mg 0,0047 mg 0,0060 mg 0,0078 mg 0,011 mg 0,014 mg 0,018 mg m·(3,5·10-7) m·(8,0·10-7) m·(2,5·10-6) m·(3,0·10-6) | m = mass |
| 5.4 |
| Pressure |
| Depression |
| - 0,9 bar to 0 bar | 0,085 mbar | Medium: gas; Measuring instruments: mechanical, electro-mechanical |
| Overpressure |
| 0 mmHg to 300 mmHg (0 kPa to 40 kPa) | 0,064 mmHg 8,5 Pa | Medium: gas Measuring instruments: blood pressure manometers | ||
| 0 bar to 1 bar | 0,085 mbar | Medium: gas, fluid; Measuring instruments: mechanical, electro-mechanical | ||
| 1 bar to 35 bar | 4 mbar | Medium: gas, fluid; Measuring instruments: mechanical, electro-mechanical, tire pressure gauges for motor vehicles | ||
| 35 bar to 250 bar | 0,3 bar | Medium: gas, fluid; Measuring instruments: mechanical, electro-mechanical | ||
| 250 bar to 600 bar | 1,7 bar | Medium: fluid; Measuring instruments: mechanical, electro-mechanical |
| Absolute pressure |
| 0,1 bar to 2 bar | 0,17 mbar | Medium: gas; Measuring instruments: mechanical, electro-mechanical |
| Evaluation of chambers |
| 0,1 bar to 5 bar | 0,15 bar | Medium: gas; Measuring instruments: mechanical, electro-mechanical |
| 5.6 |
| Torque |
| Torque wrenches |
| to 20 Nm 20 Nm to 60 Nm 60 Nm to 1000 Nm 1000 Nm to 2000 Nm 3000 Nm to 3000 Nm | 1,2 % 1,0 % 0,7 % 0,5 % 0,3 % | In laboratory and on-site |
| 5.7 |
| Acceleration |
| Measuring devices for acceleration and deceleration |
| 0 ms-2 to 10 ms-2 | 0,03 ms-2 | By comparison In laboratory and on-siteIn laboratory and on-site |
| Accelerometers |
| 100 mm/s2 to 100 m/s2 | 1,2 % | Constant force acceleration measurement transducers |
| 5.8 |
| Hardness |
| Hardness testers |
| Shore IRHD | 10 shore to 100 shore 10 IRDH to 100 IRDH | 1,0 shore 1,0 IRDH | ISO 18898 In laboratory and on-site | |
| Brinell hardness testers | 40 HB to 500 HB | 2,0 HB | ISO 6506-2 (indirect method) In laboratory and on-site | |
| Vickers hardness testers | 20 HV to 940 HV | 2,0 HV | ISO 6507-2 (indirect method) In laboratory and on-site | |
| Rockwell hardness testers | 20 HR to 85 HR | 2,0 HR | ISO 6508-2 (indirect method) In laboratory and on-site |
| 6 |
| ACOUSTICAL QUANTITIES |
| 6.1 |
| Sound Pressure level |
| Sound level meters 94 dB |
| 31,5 Hz to 12,5 kHz | 0,2 dB |
| Acoustic calibrators 94 dB |
| 31,5 Hz to 8 kHz 8 kHz to 12,5 kHz | 0,2 dB 0,25 dB |
| Microphone correction factor |
| 250 Hz | 0,2 dB | At 94 dB |
| Piston-phone 124 dB |
| 250 Hz | 0,08 dB |
| 6.1.5 |
| Acoustical calibration 94 dB |
| 1000 Hz | 0,09 dB |
| Acoustical calibration 94 dB– 114 dB step |
| 1000 Hz | 0,03 dB |
| Distortion at 94 dB, 124 dB |
| 31,5 Hz to 4 kHz | 0,25 d % | Distortion analyser |
| 6.2 | |
| Transducers of sound level pressure | 0 dB corresponds to 1 V/Pa |
| Electrical part of sound level meters |
| (40 to 134) dB (134 to 154) dB | 31,5 Hz to 12,5 kHz 31,5 Hz to 200 Hz | 0,1 dB 0,6 dB | 0dB corresponds to 1μV CMC is valid for instruments, whose microphone has all necessary characteristics | |
| (134 do 154) dB | 200 Hz do 4 kHz | 0,2 dB | ||
| (134 do 154) dB | 4 kHz do 20 kHz | 0,6 dB | ||
| Microphone sound level meters |
| -26 dB to -60 dB re, 1V 74 dB to 114 dB re, 1Pa | 0,1 dB | Microphone cartridge with or without preamplifier | ||
| Sound level meters |
| 0 dB to -60 dB re, 1pa 50 µV to 50 V | 0,1 dB re 1Pa 89 µV | Calibration with simulation without probe | ||
| 6.3 |
| Vibration transducers |
| Vibration meters |
| 0 m/s2 to 3500 m/s2 50 µV to 50 V | 0,8 % 89 µV | Calibration with simulation without probe | ||
| 7 |
| FLUID QUANTITIES |
| 7.4 |
| Volume of liquids |
| Pipettes (single-channel), dispensers and burettes |
| to 0,5 µl 0,5 µl to 2 µl 2 µl to 10 µl 10 µl to 20 µl 20 µl to 50 µl 50 µl to 100 µl 100 µl to 200 µl 200 µl to 500 µl 500 µl to 1000 µl 1 ml to 2 ml 2 ml to 5 ml 5 ml to 10 ml 10 ml to 20 ml 20 ml to 50 ml 50 ml to 100 ml 100 ml to 200 ml | 0,007 µl 0,010 µl 0,013 µl 0,014 µl 0,018 µl 0,055 µl 0,077 µl 0,15 µl 0,32 µl 0,65 µl 1,45 µl 3,10 µl 6,20 µl 14,3 µl 30,9 µl 61,7 µl | SIST EN ISO 8655 |
| Pipettes (multi-channel) |
| do 5 µl 5 µl to 10 µl 10 µl to 20 µl 20 µl to 50 µl 50 µl to 100 µl 100 µl to 200 µl 200 µl to 500 µl 500 µl to 1000 µl 1000 µl to 2000 µl | 0,036 µl 0,037 µl 0,040 µl 0,054 µl 0,09 µl 0,17 µl 0,40 µl 0,80 µl 1,61 µl | SIST EN ISO 8655 |
| Glass graduated and bulb pipettes, burettes and bottles for measuring density |
| to 2 ml 2 ml to 5 ml 5 ml to 10 ml 10 ml to 20 ml 20 ml to 25 ml 25 ml to 50 ml 50 ml to 200 ml 200 ml to 500 ml 500 ml to 1000 ml 1000 ml to 2000 ml | 1 µl 2 µl 4 µl 8 µl 10 µl 20 µl 40 µl 80 µl 110 µl 230 µl |
| Measuring cylinders and bottles |
| to 20 ml 20 ml to 50 ml 50 ml to 100 ml 100 ml to 200 ml 200 ml to 500 ml 500 ml to 1000 ml 1000 ml to 2000 ml | 0,016 ml 0,02 ml 0,03 ml 0,08 ml 0,3 ml 0,4 ml 0,6 ml |
| Volume containers |
| to 2 l 2 l to 5 l 5 l to 10 l 10 l to 25 l 25 l to 50 l 50 l to 100 l 100 l to 350 l | 1,9 ml 3,3 ml 3,8 ml 4,1 ml 12 ml 47 ml 73 ml |
| Volumetric dosing systems |
| to 20 ml 20 ml to 100 ml 100 ml to 2000 ml 2 l to 5 l 5 l to 10 l 10 l to 20 l 20 l to 50 l | 0,004 ml 0,009 ml 1,0 ml 1,1 ml 1,2 ml 5,2 ml 6,0 ml |
| Body volumen |
| to 10 cm3 10 cm3 to 100 cm3 100 cm3 to 200 cm3 | 0,1 % 0,2 % 0,3 % |
| 8 |
| OPTICAL QUANTITIES |
| 8.3 |
| Optical systems’ properties |
| Regloscopes |
| Tint meters |
| to 100 % | 0,5 % | Calibration with reference filters In laboratory and on-site |
| Regloscopes |
| 0 % to 5 % | 0,07 % | By comparison In laboratory and on-site |
| 10 |
| TEMPERATURE, HUMIDITY AND THERMO-PHYSICAL PROPERTIES |
| 10.3 |
| Liquid thermometers |
| - 40 °C to 20 °C 20 °C to 150 °C 150 °C to 280 °C | 0,04 °C 0,03 °C 0,05 °C | By comparison |
| 10.3.1 |
| Thermometers with display |
| - 80 °C to 20 °C 20 °C to 150 °C 150 °C to 280 °C 280 °C to 400 °C | 0,04 °C 0,03 °C 0,05 °C 0,25 °C | By comparison |
| 10.5 |
| Humidity |
| 10.5.2 |
| Relative humidity measuring devices |
| 66 % to 95 % 34 % to 95 % 30 % to 95 % 19 % to 95 % 11 % to 95 % 10 % to 95 % | 3,0 % | 10 °C to 20 °C 20 °C to 22 °C 22 °C to 30 °C 30 °C to 40 °C 40 °C to 50 °C 50 °C to 60 °C |
| 10.5.4 |
| Evaluation of temperature regulated chambers |
| Temperature calibration baths |
| - 70 °C to -40 °C - 40 °C to 0 °C 0 °C to 20 °C 20 °C to 100 °C 100 °C to 150 °C 150 °C to 200 °C 200 °C to 250 °C 250 °C to 300 °C | 0,048 °C 0,042 °C 0,036 °C 0,028 °C 0,033 °C 0,036 °C 0,040 °C 0,042 °C | In laboratory and on-site |
| Climate chambers: Stabilizers, dryers, vacuum dryers, thermostatic chambers with water-cooling, autoclaves, temperature chambers and rooms, refrigerators and freezers, hot baths |
| | | - 80 °C to 280 °C | 0,13 °C | In laboratory and on site |
| 66 % to 95 % 34 % to 95 % 30 % to 95 % 19 % to 95 % 11 % to 95 % 10 % to 95 % | 5,0 % | 10 °C to 20 °C 20 °C to 22 °C 22 °C to 30 °C 30 °C to 40 °C 40 °C to 50 °C 50 °C to 60 °C | ||
| 11 |
| CHEMICAL ANALYSIS; REFERENCE MATERIALS |
| 11.2.1 |
| Analytical instruments / monitors |
| 11.3 |
| pH measuring equipment |
| 0 pH to 14 pH -1999 mV to 1999 mV | 0,009 pH 0,30 mV | Calibration of measures with BNC conectors and inner impedance of 100Ω/1000MΩ - with simulation, without probe. In lab and on-site |
| Conductivity measuring equipment |
| 0,04µS/cm to 0,1mS/cm 0,1mS/cm to 800mS/cm | 0,2% 0,1% | Calibration without probe with simulation In lab and on-site |
| On site |
| Measured quantity | Range | Calibration and measuring capability CMC (k=2)* | Notes |
| 5 |
| MECHANICAL QUANTITIES |
| 5.1 |
| Force |
| Roll brake testers |
| to 0,5 kN 0,5 kN to 1,0 kN 1,0 kN to 2,0 kN 2,0 kN to 3,0 kN 3,0 kN to 6,0 kN | 1,6 % 1,4 % 1,2 % 1,0 % 0,9 % | Track for cars and mopeds | ||
| to 10 kN 10 kN to 20 kN 20 kN to 30 kN | 1,8 % 0,8 % 0,6 % | Track for trucks |
| 5.2 |
| Conventional mass |
| Etalon weights |
| 0,1 g 0,2 g 0,5 g 1 g 2 g 5 g 10 g 20 g 50 g 100 g 200 g 500 g 1 kg 2 kg 5 kg 10 kg 20 kg 500 kg | 0,15 mg 0,20 mg 0,25 mg 0,3 mg 0,4 mg 0,5 mg 0,6 mg 0,8 mg 1,0 mg 1,5 mg 3,0 mg 7,5 mg 15 mg 30 mg 75 mg 150 mg 300 mg 15 g | |||
| Objects of other optional masses |
| to 1 g 1 g to 5 g 5 g to 20 g 20 g to 50 g 50 g to 100 g 100 g to 205 g 205 g to 500 g 500 g to 1000 g 1000 g to 2100 g 2,1 kg to 5 kg 5 kg to 10 kg 10 kg to 20 kg 20 kg to 32 kg 32 kg to 400 kg | 1 mg 2 mg 3 mg 5 mg 6 mg 8 mg 15 mg 25 mg 50 mg 0,25 g 0,30 g 0,45 g 0,65 g 28 g | |||
| 5.3 |
| Scales |
| Non-automatic scales |
| to 0,005 g 0,005 g to 0,02 g 0,02 g to 0,05 g 0,02 g to 0,1 g 0,1 g to 0,2 g 0,2 g to 0,5 g 0,5 g to 1 g 1 g to 2 g 2 g to 5 g 5 g to 10 g 10 g to 20 g 20 g to 50 g 50 g to 100 g 100 g to 200 g 200 g to 25 kg 25 kg to 100 kg 100 kg to 600 kg 600 kg to 6000 kg 6 t to 37,5 t 37,5 t to 60 t 60 t to 75 t 75 t to 100 t | 0,004 mg 0,005 mg 0,006 mg 0,008 mg 0,009 mg 0,012 mg 0,015 mg 0,019 mg 0,024 mg 0,030 mg 0,039 mg 0,051 mg 0,088 mg 0,16 mg m·(8,0·10-7) m·(2,5·10-6) m·(3,0·10-6) m·(1,5·10-4) m·(2,5·10-4) m·(3,5·10-4) m·(4,0·10-4) m·(5,0·10-4) | m = mass | ||
| 5.4 |
| Pressure |
| Depression |
| - 0,9 bar to 0 bar | 0,085 mbar | Medium: gas Measuring instruments: mechanical, electro-mechanical | ||
| Overpressure |
| 0 mmHg to 300 mmHg (0 kPa to 40 kPa) | 0,064 mmHg 8,5 Pa | Medium: gas Measuring instruments: blood pressure manometers | ||
| 0 bar to 1 bar | 0,085 mbar | Medium: gas, fluid; Measuring instruments: mechanical, electro-mechanical | ||
| 1 bar to 35 bar | 4 mbar | Medium: gas, fluid; Measuring instruments: mechanical, electro-mechanical, tire pressure gauges for motor vehicles | ||
| 35 bar to 250 bar | 0,3 bar | Medium: gas, fluid; Measuring instruments: mechanical, electro-mechanical | ||
| 250 bar to 600 bar | 1,7 bar | Medium: fluid; Measuring instruments: mechanical, electro-mechanicali |
| Absolute pressure |
| 0,1 bar to 2 bar | 0,17 bar | Medium: gas; Measuring instruments: mechanical, electro-mechanical |
| Evaluation of chambers |
| 0,1 bar to 5 bar | 0,15 bar | Medium: gas; Measuring instruments: mechanical, electro-mechanical |
| 7 |
| FLUID QUANTITIES |
| 7.4 |
| Volume of liquids |
| Volumetric dosing systems |
| to 5 ml 5 ml to 10 ml 10 ml to 20 ml 20 ml to 50 ml 50 ml to 100 ml 100 ml to 200 ml 200 ml to 500 ml 500 ml to 1000 ml 1000 ml to 5000 ml 5 l to 30 l 30 l to 100 l 100 l to 200 l 200 l to 300 l 300 l to 400 l 400 l to 500 l | 0,02 ml 0,03 ml 0,04 ml 0,08 ml 0,2 ml 0,4 ml 0,8 ml 1,6 ml 3,1 ml 0,06 l 0,4 l 0,5 l 0,6 l 0,7 l 0,8 l |
| 10 |
| TEMPERATURE, HUMIDITY AND THERMO-PHYSICAL PROPERTIES |
| 10.3.1 |
| Thermometers with display |
| - 30 °C to 100 °C 100 °C to 140 °C 140 °C to 200 °C 200 °C to 300 °C 300 °C to 400 °C | 0,05 °C 0,08 °C 0,15 °C 0,20 °C 0,25 °C | By comparison |
| 10.5.4 |
| Evaluation of temperature controlled chambers |
| Temperature calibration baths |
| - 70 °C to -40 °C -40 °C to 0 °C 0 °C to 20 °C 20 °C to 100 °C 100 °C to 150 °C 150 °C to 200 °C 200 °C to 250 °C 250 °C to 300 °C | 0,048 °C 0,042 °C 0,036 °C 0,028 °C 0,033 °C 0,036 °C 0,040 °C 0,042 °C |
| Climatic chambers: SStabilizer, dryers, vacuum drayers, thermostat chambers with water cooling, autoclaves, temperature chambers and rooms, refrigrators and freezers, hot baths |
| -80 °C to 280 °C | 0,13 °C | |||
| 66 % to 95 % 34 % to 95 % 30 % to 95 % 19 % to 95 % 11 % to 95 % 10 % to 95 % | 5,0 % | 10 °C to 20 °C 20 °C to 22 °C 22 °C to 30 °C 30 °C to 40 °C 40 °C to 50 °C 50 °C to 60 °C |