Calibration is required in the process of controlling our equipment if we are to ensure the traceability of measurements.

With the help of etalon weights we verify the accuracy of our indication, eccentricity and repetability.
Force
RANGE
- to 5000 N
![]() | ![]() |
- to 20 kN
WE PERFORM
- Calibration of mechanic and electro-mechanic force transducers with method of comparison of indicating reference etalon value and indicating measuring device.
![]() | ![]() | ![]() |
SERVICE
- We perform adjustment of force transducers of all types.
WHY PERFORM CALIBRATION
- Accuracy of measuring devices is not stable time-wise.
- Change in temperature, humidity, air pressure and other adverse factors influence the accuracy of devices, as well as operator’s work.
- Regular calibration ensures accurate and precise operation of measuring devices.
- We inspect accuracy of indication with etalon weights or reference force measuring device.
- We try to eliminate established deviation of measuring devices in shortest time possible.
- In procedure of preparation for measuring device inspection, the measuring device can be set in a manner so that it can be prepared for operation, which is in accordance with its use.
Certificate, which you are provided with after each calibration, ensure measuring traceability of measuring devices with establishment and maintenance of quality systems (ISO certification), laboratory accreditation and is in accordance with requirements GMP, GLP and GCP.
CALIBRATION PERIOD
Calibration period is set in accordance with standards or recommendations for ensuring quality of measuring equipment SIST ISO 10012, ILAC-G24 / OIML D10.
DETAILED SCOPE OF ACCREDITATION
![]() | Annex to the Accreditation Certificate | LK-008 |
| Issued on | 15. junij 2011 | |
| Replaces Annex dated | 1. april 2011 | |
| Information on current accreditation status is available at: | www.sa.gov.si |
| In laboratory and on site |
| Measured quantity | Range | Best measuring capability BMC (k=2)* | Notes |
| 5 |
| MECHANICAL QUANTITIES |
| 5.1 |
| Force |
| Force transducers |
| > 1 N to 10 N 10 N to 100 N 100 N to 200 N 200 N to 400 N 400 N to 600 N 600 N to 1000 N 1000 N to 1500 N 1500 N to 2000 N 2000 N to 2500 N 2500 N to 5000 N | 0,011 N 0,11 N 0,11 N 0,11 N 0,12 N 0,13 N 1,1 N 1,1 N 1,2 N 3,5 N | By comparison In laboratory and on-site |
WHY CHOOSE US
- Slovenian accreditation constantly audit of our technical staff and equipment.
- Results of inter-laboratory comparisons show good work of the laboratory.
- We provide quick reaction time from order to service implementation.
- We strive to interfere with the work process as less as possible when providing service.
- Constant training of our staff enables us quality service.
- All of our measuring equipment has secured measuring traceability on national or international etalon.
- We inform you regularly on the date of next calibration.
- MeOL: electronic monitoring of measuring devices over the Internet.
- Trust of auditors and inspectors in our results.
- We are always available to provide information.
SALE