RANGE

  • from -0,9 bar to 600 bar
     
Ikona_slovenska_akreditacija Ikona_imenovana_oseba

WE PERFORM

  • Calibration of:
    • mechanic manometers, vacuum meters in accordance with standard SIST EN 837,
    • electro-mechanic pressure measuring devices in accordance with document EA-10/17,
    • environment pressure measuring devices.
  • Evaluation of:
    • pressure chambers.
  • Execution in our laboratory or on-site.
     
Merilnik_tlaka_v_pnevmatikah  Pregled_manometrov  Merilnik_krvnega_tlaka

WHY PERFORM CALIBRATION

  • In case of greater pressure shocks, accuracy of pressure measuring devices decreases quicker,
  • accuracy of pressure measuring devices is not stable time-wise,
  • we verify accuracy between set and actual pressure,
  • we determine hysteresis of pressure measuring devices.


Certificate, which you are provided with after each calibration, ensure measuring traceability of measuring devices with establishment and maintenance of quality systems (ISO certification), laboratory accreditation and is in accordance with requirements GMP, GLP and GCP.
 

Tlacna_komora

CALIBRATION PERIOD

Calibration period is set in accordance with standards or recommendations for ensuring quality of measuring equipment SIST ISO 10012, ILAC-G24 / OIML D10.

DETAILED SCOPE OF ACCREDITATION

Annex to the Accreditation Certificate LK-008
  Issued on 15. junij 2011
  Replaces Annex dated 1. april 2011
  Information on current accreditation status is available at: www.sa.gov.si

In laboratory
Measured quantity Range Best measuring capability BMC (k=2)* Notes
 5
MECHANICAL QUANTITIES
5.4
Pressure
 
Depression
    - 0,9 bar to 0 bar 0,085 mbar Medium: gas;
Measuring instruments: mechanical, electro-mechanical
 
Overpressure
    0 mmHg to 300 mmHg
(0 kPa to 40 kPa)
0,064 mmHg
8,5 Pa
Medium: gas
Measuring instruments: blood pressure manometers
    0 bar to 1 bar 0,085 mbar Medium: gas, fluid;
Measuring instruments: mechanical, electro-mechanical
    1 bar to 35 bar 4 mbar Medium: gas, fluid;
Measuring instruments: mechanical, electro-mechanical, tire pressure gauges for motor vehicles
    35 bar to 250 bar 0,3 bar Medium: gas, fluid;
Measuring instruments: mechanical, electro-mechanical
    250 bar to 600 bar 1,7 bar Medium: fluid;
Measuring instruments: mechanical, electro-mechanical
 
Absolute pressure
    0,1 bar to 2 bar 0,17 mbar Medium: gas;
Measuring instruments: mechanical, electro-mechanical
 
Evaluation of chambers
    0,1 bar to 5 bar 0,15 bar Medium: gas;
Measuring instruments: mechanical, electro-mechanical
On site
Measured quantity Range Best measuring capability BMC (k=2)* Notes
5
MECHANICAL QUANTITIES
5.4
Pressure
 
Depression
    - 0,9 bar to 0 bar 0,085 mbar Medium: gas
Measuring instruments: mechanical, electro-mechanical
 
Overpressure
    0 mmHg to 300 mmHg
(0 kPa to 40 kPa)
0,064 mmHg
8,5 Pa
Medium: gas
Measuring instruments: blood pressure manometers
    0 bar to 1 bar 0,085 mbar Medium: gas, fluid;
Measuring instruments: mechanical, electro-mechanical
    1 bar to 35 bar 4 mbar Medium: gas, fluid;
Measuring instruments: mechanical, electro-mechanical, tire pressure gauges for motor vehicles
    35 bar to 250 bar 0,3 bar Medium: gas, fluid;
Measuring instruments: mechanical, electro-mechanical
    250 bar to 600 bar 1,7 bar Medium: fluid;
Measuring instruments: mechanical, electro-mechanicali
 
Absolute pressure
    0,1 bar to 2 bar 0,17 bar Medium: gas;
Measuring instruments: mechanical, electro-mechanical
 
Evaluation of chambers
    0,1 bar to 5 bar 0,15 bar Medium: gas;
Measuring instruments: mechanical, electro-mechanical

WHY CHOOSE US

  • Slovenian accreditation constantly audit our technical staff and equipment,
  • we provide quick reaction time from order to service implementation,
  • constant training of our staff enables us quality service,
  • all of our measuring equipment has secured measuring traceability on national or international etalon,
  • we inform you regularly on the date of next calibration,
  • MeOL: electronic monitoring of measuring devices over the Internet,
  • trust of auditors and inspectors in our results,
  • we are always available to provide information.