Attractions

Calibration is required in the process of controlling our equipment if we are to ensure the traceability of measurements.

With the help of etalon weights we verify the accuracy of our indication, eccentricity and repetability.
Temperature
RANGE
- thermometers from -40 °C to 400 °C
- climate chambers from -40 °C to 280 °C
- calibration baths from -70 °C to 300 °C
![]() | ![]() |
WE PERFORM
- Calibration of :
- thermometers with display, liquid and radiation thermometers,
- memory thermometers – data loggers.
- Evaluation of :
- climate chamber (from -40 °C to 280 °C) – climate chambers include freezers, refrigerator, coolers, incubators, driers, climate spaces etc.,
- steam sterilizers, autoclaves – pressure evaluation is also performed
- calibration baths (from -70 °C to 300 °C),
- humidity analyzers and PCR apparatuses - cyclers,
- Execution in our laboratory (or on-site).
![]() |
USE OF MEASURING DEVICES
- In medicine for storing medication and vaccines.
- for HACCP system purposes.
- for measuring environment conditions, in accredited bodies.
- in coolers.
- in warehouses.
- in industry.
- in transport and distribution of thermal energy…
![]() |
CALIBRATION OR EVALUATION
- Thermometer calibration is performed by method of comparison with the help of calibration bath or climate chamber, with which we ensure stabile temperature condition. With calibration, measured body error is determined, which is evident from the accredited certificate on calibration, which you receive after the completed procedure. It includes results in form of a chart, which can be used as correction in your measurements.
- Evaluation of temperature regulated chambers is performed by measuring distribution of workspace temperature in the chamber for determined time period in multiple points.
WHY PERFORM CALIBRATION OR EVALUATION
- Heating and cooling of the thermometer, vibration and handling affects the sensor and therefore the thermometer accuracy.
- temperature regulated chambers, thermometers with display and liquid thermometers are not stable time-wise.
- we verify accuracy between the set and actual temperature.
- we measure uniformity of temperature distribution in the area.
Certificate, which you are provided with after each calibration, ensure measuring traceability of measuring devices with establishment and maintenance of quality systems (ISO certification), laboratory accreditation and is in accordance with requirements GMP, GLP and GCP.
CALIBRATION PERIOD
Calibration period is set in accordance with standards or recommendations for ensuring quality of measuring equipment SIST ISO 10012, ILAC-G24 / OIML D10.
DETAILED SCOPE OF ACCREDITATION
![]() | Annex to the Accreditation Certificate | LK-008 |
| Issued on | 15. junij 2011 | |
| Replaces Annex dated | 1. april 2011 | |
| Information on current accreditation status is available at: | www.sa.gov.si |
| In laboratory |
| Measured quantity | Range | Best measuring capability BMC (k=2)* | Notes |
| 10 |
| TEMPERATURE, HUMIDITY AND THERMO-PHYSICAL PROPERTIES |
| 10.3 |
| Liquid thermometers |
| - 40 °C to 20 °C 20 °C to 150 °C 150 °C to 280 °C | 0,04 °C 0,03 °C 0,05 °C | By comparison |
| 10.3.1 |
| Thermometers with display |
| - 80 °C to 20 °C 20 °C to 150 °C 150 °C to 280 °C 280 °C to 400 °C | 0,04 °C 0,03 °C 0,05 °C 0,25 °C | By comparison |
| 10.5.4 |
| Evaluation of temperature regulated chambers |
| Temperature calibration baths |
| - 70 °C to -40 °C - 40 °C to 0 °C 0 °C to 20 °C 20 °C to 100 °C 100 °C to 150 °C 150 °C to 200 °C 200 °C to 250 °C 250 °C to 300 °C | 0,048 °C 0,042 °C 0,036 °C 0,028 °C 0,033 °C 0,036 °C 0,040 °C 0,042 °C | In laboratory and on-site |
| Climate chambers: Stabilizers, dryers, vacuum dryers, thermostatic chambers with water-cooling, autoclaves, temperature chambers and rooms, refrigerators and freezers, hot baths |
| | | - 80 °C to 280 °C | 0,13 °C | In laboratory and on site |
| 66 % to 95 % 34 % to 95 % 30 % to 95 % 19 % to 95 % 11 % to 95 % 10 % to 95 % | 5,0 % | 10 °C to 20 °C 20 °C to 22 °C 22 °C to 30 °C 30 °C to 40 °C 40 °C to 50 °C 50 °C to 60 °C | ||
| On site |
| Measured quantity | Range | Best measuring capability BMC (k=2)* | Notes |
| 10 |
| TEMPERATURE, HUMIDITY AND THERMO-PHYSICAL PROPERTIES |
| 10.3.1 |
| Thermometers with display |
| 10.5.4 |
| Termometri s prikazovalnikom |
| - 30 °C to 100 °C 100 °C to 140 °C 140 °C to 200 °C 200 °C to 300 °C 300 °C to 400 °C | 0,05 °C 0,08 °C 0,15 °C 0,20 °C 0,25 °C | By comparison |
WHY CHOOSE US
- Slovenian accreditation constantly audit our technical staff and equipment.
- results of inter-laboratory comparisons show good work of the laboratory.
- we provide quick reaction time from order to service implementation.
- we strive to interfere with the work process as less as possible when providing service.
- constant training of our staff enables us quality service.
- all of our measuring equipment has secured measuring traceability on national or international etalon.
- we inform you regularly on the date of next calibration.
- MeOL: electronic monitoring of measuring devices over the Internet.
- trust of auditors and inspectors in our results.
- we are always available to provide information.