Attractions

Calibration is required in the process of controlling our equipment if we are to ensure the traceability of measurements.

With the help of etalon weights we verify the accuracy of our indication, eccentricity and repetability.

   

Temperature

Ikona_temperatura Ikona_parni_sterilizator Ikona_temperatura_izvedba_meritev Ikona_servis Ikona_ISO_IEC_17025 Ikona_certifikat_kalibracija ikona_certifikat_ovrednotenje

RANGE

  • thermometers from -40 °C to 400 °C
  • climate chambers from -40 °C to 280 °C
  • calibration baths from -70 °C to 300 °C
     
Ikona_slovenska_akreditacija_LK-008 Ikona_imenovana_oseba

WE PERFORM 

  • Calibration of :
    • thermometers with display, liquid and radiation thermometers,
    • memory thermometers – data loggers.
  • Evaluation of :
    • climate chamber (from -40 °C to 280 °C) – climate chambers include freezers, refrigerator, coolers, incubators, driers, climate spaces etc.,
    • steam sterilizers, autoclaves – pressure evaluation is also performed
    • calibration baths (from -70 °C to 300 °C),
    • humidity analyzers and PCR apparatuses - cyclers,
  • Execution in our laboratory (or on-site).
     
Olna_kopelj  Postavitev_senzorjev_v_temperaturni_komori  Spremljanje_temperature

USE OF MEASURING DEVICES

  • In medicine for storing medication and vaccines.
  • for HACCP system purposes.
  • for measuring environment conditions, in accredited bodies.
  • in coolers.
  • in warehouses.
  • in industry.
  • in transport and distribution of thermal energy…
     
Uporaba_meril_in_naprav_temperature  Temperatura_termometer  Postavitev_EBI_Senzorjev_v_komori

CALIBRATION OR EVALUATION

  • Thermometer calibration is performed by method of comparison with the help of calibration bath or climate chamber, with which we ensure stabile temperature condition. With calibration, measured body error is determined, which is evident from the accredited certificate on calibration, which you receive after the completed procedure. It includes results in form of a chart, which can be used as correction in your measurements.
  • Evaluation of temperature regulated chambers is performed by measuring distribution of workspace temperature in the chamber for determined time period in multiple points.

WHY PERFORM CALIBRATION OR EVALUATION

  • Heating and cooling of the thermometer, vibration and handling affects the sensor and therefore the thermometer accuracy.
  • temperature regulated chambers, thermometers with display and liquid thermometers are not stable time-wise.
  • we verify accuracy between the set and actual temperature.
  • we measure uniformity of temperature distribution in the area.

Certificate, which you are provided with after each calibration, ensure measuring traceability of measuring devices with establishment and maintenance of quality systems (ISO certification), laboratory accreditation and is in accordance with requirements GMP, GLP and GCP.

CALIBRATION PERIOD

Calibration period is set in accordance with standards or recommendations for ensuring quality of measuring equipment SIST ISO 10012, ILAC-G24 / OIML D10. 

DETAILED SCOPE OF ACCREDITATION

Annex to the Accreditation Certificate LK-008
  Issued on 15. junij 2011
  Replaces Annex dated 1. april 2011
  Information on current accreditation status is available at: www.sa.gov.si

In laboratory
Measured quantity Range Best measuring capability BMC (k=2)* Notes
10
TEMPERATURE, HUMIDITY AND THERMO-PHYSICAL PROPERTIES
10.3
Liquid thermometers
    - 40 °C to 20 °C
20 °C to 150 °C
150 °C to 280 °C
0,04 °C
0,03 °C
0,05 °C
By comparison
10.3.1
Thermometers with display
    - 80 °C to 20 °C
20 °C to 150 °C
150 °C to 280 °C
280 °C to 400 °C
0,04 °C
0,03 °C
0,05 °C
0,25 °C
By comparison
10.5.4
Evaluation of temperature regulated chambers
 
Temperature calibration baths
    - 70 °C to -40 °C
- 40 °C to 0 °C
0 °C to 20 °C
20 °C to 100 °C
100 °C to 150 °C
150 °C to 200 °C
200 °C to 250 °C
250 °C to 300 °C
0,048 °C
0,042 °C
0,036 °C
0,028 °C
0,033 °C
0,036 °C
0,040 °C
0,042 °C
In laboratory and on-site
 
Climate chambers:
Stabilizers, dryers, vacuum dryers, thermostatic chambers with water-cooling, autoclaves, temperature chambers and rooms, refrigerators and freezers, hot baths
 
 
 
 
- 80 °C to 280 °C 0,13 °C In laboratory and on site
66 % to 95 %
34 % to 95 %
30 % to 95 %
19 % to 95 %
11 % to 95 %
10 % to 95 %
5,0 % 10 °C to 20 °C
20 °C to 22 °C
22 °C to 30 °C
30 °C to 40 °C
40 °C to 50 °C
50 °C to 60 °C

On site
Measured quantity Range Best measuring capability BMC (k=2)* Notes
10
TEMPERATURE, HUMIDITY AND THERMO-PHYSICAL PROPERTIES
10.3.1
Thermometers with display
10.5.4
Termometri s prikazovalnikom
    - 30 °C to 100 °C
100 °C to 140 °C
140 °C to 200 °C
200 °C to 300 °C
300 °C to 400 °C
0,05 °C
0,08 °C
0,15 °C
0,20 °C
0,25 °C
By comparison

WHY CHOOSE US

  • Slovenian accreditation constantly audit our technical staff and equipment.
  • results of inter-laboratory comparisons show good work of the laboratory.
  • we provide quick reaction time from order to service implementation.
  • we strive to interfere with the work process as less as possible when providing service.
  • constant training of our staff enables us quality service.
  • all of our measuring equipment has secured measuring traceability on national or international etalon.
  • we inform you regularly on the date of next calibration.
  • MeOL: electronic monitoring of measuring devices over the Internet.
  • trust of auditors and inspectors in our results.
  • we are always available to provide information.

SALE

Logotip_Dostmann

TRAINING

Ikona_izobrazevanje