Steam sterilizers

Inspection is performed in accordance with standard SIST EN 285 for large sterilizers and SIST EN 13060 for small sterilizers.

Ikona_temperatura Ikona_certifikat_kontrola Ikona_certifikat_kalibracija

STERILIZER THERMOMETER CALIBRATION

Thermometer calibration is performed with comparative method with the help of calibration bath or thermoblock, with which we ensure stable temperature. In sterilization phase, a comparison of temperature probe of autoclave will be made with the value of nearby control probe.

Ikona_standardi Kalibracija_termometra_sterilizarorja
Ikona_temperatura Ikona_certifikat_kontrola Ikona_certifikat_kalibracija

STERILIZER MANOMETER CALIBRATION

Calibration is performed in accordance with standard SIST EN 837 and guideline EURAMETIcg-17lv.O1:

  • pressure depression – mechanic and electro-mechanic manometers,
  • overpressure - mechanic and electro-mechanic manometers,
  • absolute pressure – air and absolute pressure measuring instruments.

We compare pressure values on the manometer of the sterilizer and values, which will be recorded with reference instruments at value of absolute (environmental) pressure, in time of vacuum and sterilization.
 

Ikona_standardi Kalibracija_manometra_avtoklava
Ikona_cas Ikona_certifikat_kontrola Ikona_certifikat_kalibracija

STERILIZATION TIME CONTROL

We compare time on the reference instrument and on the readout from the sterilizer printer.
 

Ikona_standardi   Kontrola_casa_sterilizacije
Ikona_tlak Ikona_certifikat_kontrola

STERILIZER TIGHTNESS CONTROL

  • Appropriate sterilizer tightness is determined.
  • We establish change of pressure in vacuum with a reference pressure measuring instrument in fixed time period.
     
Ikona_standardi Kontrola_tesnosti_sterilizatorjaPokrov_avtoklava
Ikona_avtoklav Ikona_certifikat_kontrola

BOWIE & DICK TEST

Bowie & Bowie & Dick test program serves as a functional test of eliminating air and of steam permeability in package. We put the test package in the chamber, choose the program and engage the sterilizer. The test is successful, if the test package colors evenly along the entire surface.
 

Ikona_standardi Bowie_&-_Dick_test
Ikona_avtoklav Ikona_certifikat_kontrola

BIOLOGICAL TESTING

  • Biological testing is done in the framework of heat penetration.
  • We put a biological indicator next to each probe of the reference sensor.
     
Ikona_standardi Biološki_indikator
Ikona_avtoklav Ikona_certifikat_kontrola

PROGRAM FUNCTIONS CONTROL

Program functions control is performed during complete control:

  • We examine sterilizer operation and process of work programs.
  • We ensure that the main penetration control tests are performed with properly working equipment.
     
Ikona_standardi Kontrola_programskih_funkcij
Ikona_avtoklav Ikona_certifikat_kontrola

HEAT DISTRIBUTION CONTROL

Control is performed in accordance with standard SIST EN 60068.

  • Heat distribution in the sterilizer is executed in an empty chamber. The purpose of heat distribution control is to determine the coldest and the hottest point inside the sterilizer.
     
Ikona_standardi Kontrola_distribucije_toplote
Ikona_avtoklav Ikona_certifikat_kontrola

HEAT PENETRATION CONTROL

Control is executed in accordance with standard SIST EN 60068.

  • The purpose of testing is to determine the coldest areas inside a certain feed, in the area of individual units, as well as in the entire chamber. FO value is calculated with control.
     
Ikona_standardi Kontrola_penetracije_toplote

SERVICE

  • We can perform service for all kinds of sterilizers.
  • We provide spare parts for most manufacturers of sterilizers.

INSPECTION PERIOD

The inspection process of measuring devices serves as a technical part of accreditation. The inspection can thus be carried out in connection with the accreditation period which is determined by the law.
Ohterwise the period is determined according to the standards or recommendations for ensuring the quality of measuring devices SIST ISO 10012, ILAC-G24 / OIML D10.

DETAILED SCOPE OF ACCREDITATION RANGE

 No.
Inspection field
(product, product design, service, process, plant)
Inspection type
Methods and procedures,
normative documents
Notes
14. Inspection of steam sterilizers, autoclaves -initial inspection
-subsequent inspection (periodic)
-extraordinary inspection
SIST EN 285:2006 Srerilization-Steam sterilizers - Large sterilizers SIST EN ISO 17665-1:2006 Sterilization of health care Inspection in the premises of the inspection body and on site.
18. Protection equipment:
- safety valves
- inspection of adjustment and functionality Intemal procedure ML10N80, issue 3 Inspection is petformed for safety valves which form part of pressure equipment specified in items 15th, 16th and 17th of this table.

Inspection only on place of instrument use.

WHY CHOOSE US

  • Slovenian accreditation constantly audit of our technical staff and equipment,
  • in the process of checking and cleaning is carried out to verify the functionality criteria, and its accuracy in accordance with the requirements of standards and recommendations,
  • we provide quick reaction time from order to service implementation,
  • constant training of our staff enables us quality service,
  • all of our measuring equipment has secured measuring traceability on national or international etalon,
  • we inform you regularly on the date of next calibration,
  • MeOL: electronic monitoring of measuring devices over the Internet,
  • trust of auditors and inspectors in our results.